AudioDir 2017, 1, 8

Article

Faire étalonner son audiomètre. Quand, pourquoi et comment ?

Première partie : Principes et aspects théoriques

Have one’s audiometer calibrated. When, Why, How?

First part: Principles and theoretical aspects

Christian Meyer-Bisch

Point Neuve Point Bottes 40 in Bottes in Neuve 40 6q6wHER

Médecin, consultant en audiologie, expert à l’Afnor et à l’ISO.

Auteur de correspondance : christian.meyer-bisch@wanadoo.fr

Lo Converse Femmes 3 Bleu 161422c 8 Ox Uk CtasRésumé

L’étalonnage d’un audiomètre doit être contrôlé régulièrement ainsi que cela se fait pour la plupart des instruments de mesure. Le vieillissement de certains composants de l’appareil lui-même ou de ses transducteurs, son transport éventuel ou, tout simplement, l’infl uence de son environnement sont à tout moment susceptibles de modifi er les caractéristiques de la chaîne de mesure. Pour harmoniser ces contrôles, on se réfère à des normes internationales, qui peuvent être matérielles (CEI) ou méthodologiques (ISO). Ces normes sont adoptées dans la quasi-totalité des pays du monde et sont réexaminées au plus tard tous les cinq ans. Certains contrôles de routine sont à effectuer avant chaque utilisation, mais les contrôles d’étalonnage qui demandent l’intervention de techniciens spécialisés devraient être réalisés chaque année. Dans une première partie, nous aborderons les aspects techniques, et dans la seconde, les procédures mises en oeuvre.

Abstract

The calibration of an audiometer must be controlled regularly like that is done for most measuring instruments. The ageing of certain components of the device itself or its transducers, its possible transport or, quite simply, the infl uence of its environment are constantly likely to modify the characteristics of the measuring equipment. To harmonize these controls, one refers to international standards, which can be material (CIS) or methodological (ISO). These standards are adopted in the near total of the countries of the world and are re-examined at the latest every fi ve years. Certain routine inspections are to be carried out before each use, but controls of calibration which ask for the intervention of specialized technicians should be realized each year. In a fi rst part, we will approach the technical sides, and in the second, the procedures put in work.

Mots clés : Audiomètre / étalonnage / contrôles périodiques

Key words: Audiometer / calibration / periodic controls

© EDP Santé, 2017